Issue 01/2006 Art. No. 7001-0026
Operating Manual
APT.line
®
BD (E2)
Incubators with natural convection
APT.line
®
ED (E2)
Heating/drying ovens with natural convection
APT.line
®
FD (E2)
Heating/drying ovens with forced convection
with R3 Controller
BINDER GmbH
Address Post office box 102
D-78502 Tuttlingen
Tel. +49 7462 2005 0
Fax +49 7462 2005 93 0
Internet http://www.binder-world.com
Service Hotline +49 7462 2005 555
Service Fax +49 7462 2005 93 555
Service E-Mail [email protected]
Service Hotline USA +1 866 816 8191
Service Hotline Asia Pacific +603 6204 2855
Vista de pagina 0
1 2 3 4 5 6 ... 48 49

Indice de contenidos

Pagina 1 - Operating Manual

Issue 01/2006 Art. No. 7001-0026 Operating Manual APT.line® BD (E2) Incubators with natural convection APT.line® ED (E2) Heating

Pagina 2 - EG - KONFORMITÄTSERKLÄRUNG

BD / ED / FD 01/2006 page 10/49 Dear customer, For the proper operation of the incubators BD and heating/drying ovens ED and FD, it is necessar

Pagina 3

BD / ED / FD 01/2006 page 11/49 CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in moderate or minor (reve

Pagina 4

BD / ED / FD 01/2006 page 12/49 Information to be observed in order to ensure optimum function of the product. Word message panel structure T

Pagina 5

BD / ED / FD 01/2006 page 13/49 1.4 Type plate Figure 2: Type plate (example: FD 115 regular unit) Indications of the type plate Information N

Pagina 6

BD / ED / FD 01/2006 page 14/49 The incubators BD and heating/drying ovens ED and FD must NOT be operated in hazardous locations. DANGER Explos

Pagina 7

BD / ED / FD 01/2006 page 15/49 1.6 Intended use 1.6.1 Incubators BD The incubators BD are designed for exact temperation of harmless materials.

Pagina 8 - Content

BD / ED / FD 01/2006 page 16/49 5. Should the defect occur within 12 months after delivery, we shall take over all costs necessary for the repa

Pagina 9

BD / ED / FD 01/2006 page 17/49 3. Unit description BINDER incubators BD and heating/drying ovens ED and FD are equipped with an electronic PID-c

Pagina 10 - WARNING

BD / ED / FD 01/2006 page 18/49 4. Scope of delivery, transportation, storage, and installation 4.1 Unpacking, and checking equipment and scope

Pagina 11 - CAUTION

BD / ED / FD 01/2006 page 19/49 4.3 Storage Intermediate storage of the unit is possible in a closed and dry room. Respect the advice for tempora

Pagina 12 - Word message panel structure

BD / ED / FD 01/2006 page 2/49 EG - KONFORMITÄTSERKLÄRUNG EC - DECLARATION OF CONFORMITY CE - DECLARATION DE CONFORMITE Anbieter / Supplier

Pagina 13

BD / ED / FD 01/2006 page 20/49 5. Installation of the equipment 5.1 Electrical connection • BD all volumes, ED up to volume 240, FD all volum

Pagina 14

BD / ED / FD 01/2006 page 21/49 CAUTION The exhaust duct will become hot during operation. Danger of burning. ∅ Do NOT touch the exhaust duct

Pagina 15 - 2. Warranty

BD / ED / FD 01/2006 page 22/49 7. Controller setting 7.1 Display / entry of temperature set-point (without ramp function) Controller setting is

Pagina 16 - Date: November 2003

BD / ED / FD 01/2006 page 23/49 The display shows alternately “SP” and in the entry level the previous temperature set-point (example: 60°C):

Pagina 17 - 3. Unit description

BD / ED / FD 01/2006 page 24/49 7.3.1 Switching between Continuous operation and Timer operation Press the time management button . The contro

Pagina 18

BD / ED / FD 01/2006 page 25/49 7.3.3 Timer operation: Setting the tempering time 1. Press the time management button . The controller indica

Pagina 19

BD / ED / FD 01/2006 page 26/49 7.4 User level settings By pressing down button in normal display (actual value display) for 5 sec, you enter

Pagina 20

BD / ED / FD 01/2006 page 27/49 7.4.2 Enter a temperature ramp Temperature ramps can be programmed in order to extend heating up times. This may

Pagina 21 - 6. Start up

BD / ED / FD 01/2006 page 28/49 1. Press down button for approx. 5 seconds. The display alternately shows “unit” and in the entry level the te

Pagina 22 - 7. Controller setting

BD / ED / FD 01/2006 page 29/49 8. Safety devices 8.1 Temperature safety device class 2 (DIN 12880) ED, FD The temperature safety device class

Pagina 23

BD / ED / FD 01/2006 page 3/49 Die oben beschriebenen Produkte sind konform mit folgenden EG-Richtlinien: The products described above are in co

Pagina 24 - 7.3.2 Continuous operation

BD / ED / FD 01/2006 page 30/49 1. Turn the control knob (7) of the safety device using a coin to its end-stop (position 10) (unit protection).

Pagina 25

BD / ED / FD 01/2006 page 31/49 If the safety device has assumed the regulation function (identifiable by the red alarm lamp (7a) lighting up an

Pagina 26 - User level overview:

BD / ED / FD 01/2006 page 32/49 9.2 Communication software APT-COM® 3 DataControlSystem (BD, option ED) The oven is regularly equipped with a se

Pagina 27 - 7.4.3 Chamber addressing

BD / ED / FD 01/2006 page 33/49 9.5 Water protected internal socket (option BD) The internal socket is splash proof. IP system of protection 65

Pagina 28

BD / ED / FD 01/2006 page 34/49 We recommend entering a maintenance agreement. Please consult the BINDER service department. BINDER telephone ho

Pagina 29 - 8. Safety devices

BD / ED / FD 01/2006 page 35/49 (3) If necessary, have strongly contaminated inner chamber parts removed by an engineer for cleaning, or have t

Pagina 30

BD / ED / FD 01/2006 page 36/49 11.2 Decommissioning Switch off ED units sizes 400 and 720 at the main switch (10). Disconnect the unit from the

Pagina 31 - 9. Options

BD / ED / FD 01/2006 page 37/49 Do NOT place samples outside this usable volume. Do NOT load this volume more than half to enable sufficient ai

Pagina 32 - 2 3

BD / ED / FD 01/2006 page 38/49 Unit size BD 23 BD 53 BD 115 BD 240 BD 400 BD 720Temperature data Temperature range, 5°C above ambient up t

Pagina 33

BD / ED / FD 01/2006 page 39/49 12.5 Technical data series ED Unit size ED 23 ED 53 ED 115 ED 240 ED 400 ED 720Exterior dimensions Width m

Pagina 34

BD / ED / FD 01/2006 page 4/49 EG - KONFORMITÄTSERKLÄRUNG EC - DECLARATION OF CONFORMITY CE - DECLARATION DE CONFORMITE Anbieter / Supplier

Pagina 35 - 11. Disposal

BD / ED / FD 01/2006 page 40/49 Unit size ED 23 ED 53 ED 115 ED 240 ED 400 ED 720Electrical data IP system of protection acc. to EN 60529

Pagina 36 - 12. Technical description

BD / ED / FD 01/2006 page 41/49 Unit size FD 23 FD 53 FD 115 FD 240 Interior dimensions Width mm inch 222 8.74 400 15.75 600 23.62 800 31.50

Pagina 37

BD / ED / FD 01/2006 page 42/49 part 2, respecting the recommended wall clearances of 10 % of the height, width and depth of the inner chamber. T

Pagina 38

BD / ED / FD 01/2006 page 43/49 12.8 Equipment and Options Series ED To operate the heating/drying oven ED, use only original BINDER accessori

Pagina 39

BD / ED / FD 01/2006 page 44/49 12.9 Equipment and Options Series FD To operate the heating/drying oven FD, use only original BINDER accessori

Pagina 40

BD / ED / FD 01/2006 page 45/49 12.10 Spare parts The BINDER GmbH is responsible for safety-related unit properties only if skilled electrician

Pagina 41

BD / ED / FD 01/2006 page 46/49 Unit size 23 53 115 240 400 720 Description Art. No. Unit fuse 5x20mm 250V 10A semi time lag (M) 5006-0012 5

Pagina 42

BD / ED / FD 01/2006 page 47/49 13. Contamination clearance certificate Unbedenklichkeitsbescheinigung Declaration of harmlessness with regard

Pagina 43

BD / ED / FD 01/2006 page 48/49 3.3 Measures to be taken in case of skin contact or release into the atmosphere / Maßnahmen bei Personenkontakt

Pagina 44

BD / ED / FD 01/2006 page 49/49 We herewith declare that the following measures have been taken / Wir erklären, dass folgende Maßnahmen getroffe

Pagina 45 - 12.10 Spare parts

BD / ED / FD 01/2006 page 5/49 Die oben beschriebenen Produkte sind konform mit folgenden EG-Richtlinien: The products described above are in c

Pagina 46 - -- -- -- 5006-0013 --

BD / ED / FD 01/2006 page 6/49 EG - KONFORMITÄTSERKLÄRUNG EC - DECLARATION OF CONFORMITY CE - DECLARATION DE CONFORMITE Anbieter / Supplier

Pagina 47

BD / ED / FD 01/2006 page 7/49 Die oben beschriebenen Produkte sind konform mit folgenden EG-Richtlinien: The products described above are in c

Pagina 48

BD / ED / FD 01/2006 page 8/49 Content 1. SAFETY.......

Pagina 49

BD / ED / FD 01/2006 page 9/49 10. MAINTENANCE, CLEANING, AND SERVICE .............. 33 10.1 Maintenance